Products Oryx8 Oryx4 OryxNano Search Contact Us

 

Douglas Instruments
Version 2.0
Screening - using microbatch
Step by Step Instructions for Beginners

Hardware Preparation

  1. Clean a Nunc HLA plate with compressed air.

  2. Use a ground glass syringe or pipette to dispense 6 ml decane into a Nunc HLA plate. (Decane is used because its low viscosity encourages the droplet to adhere to the bottom of the well.  Paraffin oil can be used instead, but it sometimes gives less reproducible dispensing.)  Use a syringe or tip the plate to remove 4.5 ml. of excess decane, leaving decane in the wells.

  3. Place the Nunc HLA plate in the left side of the stainless steel frame on the Plate Loader. Place 7 Nunc Strips on the right side, with water for rinsing the Microtip in the first strip, and your screening solutions in the remaining six, as shown below.

dosscset.gif (13461 bytes)

  1. Connect a 2-channel Microtip to the first two channels (green and red).  Place the tip in the 2-channel "collet" (holder) on the arm of the Plate Loader.

  2. Fill the ground glass syringes of the upper valves with degassed pure water and replace them.

Software Preparation

  1. Switch on the computer and the MCC control unit.   Start Windows Explorer.  Navigate to C:\Crystallization Experiments\Screens\.   Double click on Microbatch Screens.xpp (the extension may not be visible).

  2. The program Wasprun will load and the screen will show a representation of the plates on the Plate Loader. If desired, click on Edit to change the number of wells dispensed, the volume of droplets, the percentage of protein in droplets etc. Note that V indicates a viscous solution, N indicates a non-viscous solution and R indicates a rinse well.

  3. When your experiment is ready, press or click on F2 to save it.

  4. Select Run, Dispense.  

  5. You will now move to a separate program walled Wasp.   To the left of the start button is a button with an icon showing the dispensing chassis.  Click on this button which will in turn load the program Front Panel, which is used for controlling the robotics.

  6. If you suspect that any motor positions of the syringe drivers or the Plate Loader are incorrect, select MCC, Rezero All Motors and follow instructions. 

Debubbling and Setting the Height of the Microtip.

  1. At the beginning of each day the system will require debubbling. Place a small bottle or vial under the Microtip in the Z-arm, select Syringes, Debubble, and follow instructions.  After debubbling syringes for about 10 seconds click on the red stop button or press the space bar. Any air bubbles that were present at the top of the motorized syringes should have passed out into the connecting tubing.

  2. Now follow instructions and remove the air bubbles from the connecting tubing as follows:
    a.  Remove the PTFE tubing from the needles of debubbled motorized syringes.
    b.  Expel water and air bubbles from the tubing using the ground glass syringe.
    c.  Reconnect the tubing carefully, ensuring no air bubbles re-enter.

  3. Occasionally there may be bubbles between the upper and lower valves.  Turn the top valves to the flushopttip.gif (1895 bytes) position (|-) and flush the bubbles out through the microtip with the ground glass syringes.

  4. If you suspect that the 5-channel Microtip is not set to the correct height select Plate Loader, Install Tip.  The arm(s) will move to its lowest position. Follow the instructions on the screen: (1) adjust the height of the Microtip until it is just touching the table. (2) Tighten the collet. (3) Mark the height by moving the two o-rings down to the top of the collet.

Running the Experiment

  1. Press Alt/Tab to return to Wasp.

  2. The system will now check the positions of the motorized syringes. If necessary, you will be instructed to turn valves and the syringes will be moved in order to carry out the experiment.

  3. A page of configuration information and instructions will now appear. Note that all air bubbles must be removed from all tubes of the first two channels including the microtip.  Remember to remove any droplet on the end of the microtip to prevent this from being sucked in later on. 

  4. Air will now be loaded into the microtip to separate the protein from water already in the microtip.  Some proteins may become denatured which may cause the air bubble to break up.  If you encounter such problems place the tip in an inert gas such as argon or nitrogen.  Click OK to begin the experiment.

  5. After loading air, you will be asked to provide protein. Remove the tip from the collet (the collet remains on the Z-arm) and dip it into your protein sample, avoiding any precipitate. Wipe the tip, return it to the collet on the Z-arm and do up the thumb screw.

  6. Click OK to carry out the experiment automatically. When it is finished, "top up" the HLA plate with 5 ml 50:50 silicone and paraffin mixture (or 100% paraffin to reduce evaporation). Place the plate in an incubator at the desired temperature.

  7. Quit Front Panel before turning off your computer so that the motor positions are stored for the next use of the equipment.